top of page

Temporary pacing

Indications

Class

Class I

•Is recommended

•Should be performed

​

Class IIA

•Can be effective/beneficial

​

Class IIB

•May/might be reasonable

•Effectiveness is unknown/uncertain/not well established

​

Class III

•Not recommended – ineffective/not beneficial

•Potentially harmful – associated with excess mortality

​

Bradycardia

SND (prolonged sinus pauses/bradycardia related ventricular arrhythmias/severe symptomatic SBc

•I    Acute MI with SND and symptomatic or hemo compromise

•IIA  persistent hemodynamically unstable SND (until PPM)

•IIB  severe symptoms or hemodynamic compromise (until PPM)

•III  minimal/infrequent symptoms with no hemo compromise

​

VN block (2nd type II AV block, 3rd AV block, standstill due to AVB)

•I  Acute MI and AVB with symptomatic or hemo compromise

•IIA  2nd/3rd AV block with symptoms or hemo compromise

•IIB  2nd/3rd AV block with hemo compromise (until PPM)

Procedural

•TAVI/BAV – rapid pacing

image.png

Other

•Dependent patients for generator changes

•Dependent patients for device extraction

image.png

Complications

•Pericardial Tamponade  

•Pneumothorax (air between lung & chest –causing collapsed lung) 

•Non-pericardial bleeding 

•Infection/infective endocarditis

•Diaphragmatic stimulation

•Inadvertent induction of cardiac arrhythmia (VT/VF)

•Asynchrony of chambers (VVI mode)

•Venous thrombosis

image.png

Equipment

•Pacing Box

•Pacing wire  

-Swan Ganz balloon tip + red Edwards adaptors

-Pacel (comes with own adaptors)

•Junction connector

image.png
image.png

Insertion sites

image.png
image.png

Pacing box

image.png

Pacing box 5388

1.  Pace/Sense LEDs

2.  Lock/Unlock Key

3.  Lock Indicators

4.  Rate Dial

5.  Atrial Output Dial

6.  Ventricular Output Dial

7.  Menu Parameter Dial

8.  Parameter Selection Key

9.  Menu Selection Key

10.  Pause Key

11.  Power On Key

12.  Power Off Key

13.  Emergency/Asynchronous Pacing Key

14.  Lower Screen

15.  Ventricular Output Graphics

16.  Atrial Output Graphics

17.  Upper Screen

18.  Rate Graphics

19.  Setup Indicators

20.  DDI Indicator

21.  Low Battery Indicator

22.  Setup Labels

image.png

Pacing box 5392

image.png
image.png
image.png
image.png
image.png
image.png
image.png

Electrode position

image.png
image.png

Pacing Modes

Threshold testing

•Threshold is the minimum amount of energy required to capture the heart (result in an evoked response- paced complex on ECG)

​

•It is necessary to know the threshold to ensure appropriate safety margin is programmed for the output (at least double)

image.png

How to perform

1.  Set RATE at least 10 ppm above patient’s intrinsic rate.

​

2.  Decrease OUTPUT: Slowly turn OUTPUT dial counterclockwise until ECG shows loss of capture (on ECG).

​

3.  The last output where capture was seen is labelled as the threshold (units mA).

​

4.  Set OUTPUT to a value 2 to 3 times greater than the stimulation threshold value.
This provides at least a 2:1 safety margin.

​

5.  Restore RATE to previous value.

​

Sensing testing

•Appropriate sensing needs to be determined, to ensure appropriate pacing responses occur

​

•The size of the intrinsic R wave/QRS complex should be determined to ensure appropriate sensitivity margin is programmed (usually ~half of R wave)

image.png

How to perform

1.Set RATE at least 10 ppm below patient’s intrinsic rate and set the OUPUT at sub threshold (below threshold-0.1mV). You will see that the pacing box is flashing sensing in accordance with each intrinsic beat

image.png

2.  Increase SENSITIVITY: Slowly turn sensitivity dial (in menu section) counterclockwise until flashing in the sensing window no longer occurs. Pacing flashing will then occur at the set pacing rate. Competitive pacing will not occur as the output is set below threshold.

image.png

3.  The sensitivity value where sensing was last seen is labelled as the sensing threshold or size of the R wave.

4.  Set SENSITVITY to a value around half of the sensing threshold value (units mV)
This provides at least a 2:1 safety margin.

5.  Restore RATE and OUTPUT to previous values.

Video on testing threshold and sensing testing in a single chamber temp system- > click below link

Troubleshooting

Identify the problem

•Loss of capture

•Loss of output

•Oversensing

•Undersensing

​

Identify Solution and carry out protective procedures

Loss of capture

Possible Causes

•Increased thresold

•Lead moved/dislodged

•Battery depletion

•Fracture lead

•Faulty cable connections

​

Correct the problem

•Increases output

•Reposition lead

•Replace battery

•Replace lead

•Check/change connections

image.png

Loss of output

Possible Causes

•Oversensing

•Pacemaker OFF

•Battery depletion

•Fractured lead

•Faulty cable connections

​

Correct the problem

•Decrease sensitivity

•Turn pacing box ON

•Replace battery

•Replace lead

•Check/change connections

image.png

Oversensing

Possible Causes

•T wave oversening

•Environmental interference

•Fractured lead

•Faulty cable connections

​

Correct the problem

•Increase sensitivity

•Eliminate interference

•Replace lead

•Check/change connections

image.png

Undersensing

Possible Causes

•Decreased QRS size

•Lead moved/dislodged

•Battery depletion

•Fracture lead

•Faulty cable connections

​

Correct the problem

•Increase sensitivity

•Reposition lead

•Replace battery

•Replace lead

•Check/change connections

image.png
bottom of page