Boston Scientific
Presenting rhythm & percentage paced
Presenting rhythm will be displayed at the top of the screen and percentage paced will be displayed in the summary screen

Battery
Battery data is displayed on the summary screen.
​
Selecting the battery icon will lead to more detail.

Leads
To view lead trends, select the leads tab off the summary screen.


To update lead impedance, select the pace impedance arrow under tests.

Sensing and ULR
To perform a sensing test, select appropriate mode and rate and press the intrinsic amplitude arrow under tests.

Thresholds
To perform a threshold test, select the magnifier glass for atrial of V threshold under tests.
​
Choose the appropriate mode and rate and hit start.


Observations
To view diagnostic information, select the events tab along the top.
​
Trend view will show diagnostics like events, heart rate histos, etc over time.
​
To view arrhythmia episodes select arrhythmia logbook and click magnifier glass to view episode EGM.
​
Under patient diagnostics episode totals, heart rate histos and heart rate variability can be viewed by selecting the magnifier glass icon.







Program
To view or make programming changes select settings along the top bar, settings summary and they choose therapy under atrial tachy or settings under brady.

To save PDF select Data along the bottom tabs, ensure all appropriate reports are selected and select save.