Microport
Microport: Welcome
Summary
Single chamber
•hysteresis
Dual chamber
•AAI-SafeR
Microport: About
Hysteresis
Accessed from Basic Parameters; hysteresis
•Hysteresis: 0%, 5%, 10%, 20%, 35%
Microport: About

Microport: About
Sorin – AAI-SafeR
The device will switch to DDD under 4 circumstances
​
1.The device will switch to DDD when there are 2 consecutive blocked sensed or paced atrial events (non programmable)
Microport: About

Microport: About
2. the device senses 3 blocked sensed or paced atrial events in the last 12 ventricular cycles; this parameter is non programmable
Microport: About

Microport: About
4. a ventricular pause is longer than the programmed value (parameter: Max Pause); this parameter can be programmed to 2, 3 or 4 seconds
Microport: About

Microport: About
4. a ventricular pause is longer than the programmed value (parameter: Max Pause); this parameter can be programmed to 2, 3 or 4 seconds

Microport: About
The device switches to AAI from DDD under 2 circumstances
​
1.when it senses 12 consecutive spontaneous ventricular events

Microport: About
2. automatically every 100 paced ventricular cycles

Microport: About
Accessed from Basic Parameters; hysteresis
•AVB I switch: Exercise, rest+exercise
•Long PR min and long PR max:
-programmable Long PR: 250 – 300 – 350 – 400 – 450 ms
-programmable min PR: 200 – 250 – 300 – 350 – 400 – 450 ms
•Pause V max: 2 – 3 – 4 seconds
Microport: About