Microport: About

Hysteresis

Accessed from Basic Parameters; hysteresis

•Hysteresis:  0%, 5%, 10%, 20%, 35%

Microport: About
mp 1.png
Microport: About

Sorin – AAI-SafeR

The device will switch to DDD under 4 circumstances

​

1.The device will switch to DDD when there are 2 consecutive blocked sensed or paced atrial events (non programmable)

Microport: About
mp2.jpg
Microport: About

2. the device senses 3 blocked sensed or paced atrial events in the last 12 ventricular cycles; this parameter is non programmable

Microport: About
mp3.jpg
Microport: About

4. a ventricular pause is longer than the programmed value (parameter: Max Pause); this parameter can be programmed to 2, 3 or 4 seconds

Microport: About
mp4.jpg
Microport: About

4. a ventricular pause is longer than the programmed value (parameter: Max Pause); this parameter can be programmed to 2, 3 or 4 seconds

mp5.jpg
Microport: About

The device switches to AAI from DDD under 2 circumstances

​

1.when it senses 12 consecutive spontaneous ventricular events

mp6.jpg
Microport: About

2. automatically every 100 paced ventricular cycles

mp7.jpg
Microport: About

Accessed from Basic Parameters; hysteresis

•AVB I switch: Exercise, rest+exercise

•Long PR min and long PR max:

-programmable Long PR: 250 – 300 – 350 – 400 – 450 ms

-programmable min PR: 200 – 250 – 300 – 350 – 400 – 450 ms

•Pause V max: 2 – 3 – 4 seconds

Microport: About
mp8.png
Microport: About
mp9.png
Microport: About