top of page

Biotronik

Biotronik: Welcome

Summary

Single and dual chamber:

•Hysteresis with AV search for dual chamber


Dual chamber:

•AV hysteresis and IRS+

•Ventricular pace suppression

Biotronik: About

AV search

Positive AV hysteresis. After 180 Vp beats -Opens up the AV delay by x ms for 5 cycles (AV scan cycle).


When an intrinsic R wave is detected, the AV delay is extended x ms for 5 cycles  (AV repetitive AV hysteresis).

Biotronik: About
btk search.png
Biotronik: About
btk search2.png
Biotronik: About

Accessed from Parameters; Bradycardia; Dynamic AV delay; AV hysteresis

• AV hysteresis (ppm): OFF; Negative, low 250ms; medium 285ms; high 325ms; IRSplus 400ms

•AV repetitive hysteresis (cycles): OFF; 1 ... (1) ... 10

•AV scan hysteresis (cycles): OFF; 1 ... (1) ... 10

​

btk search3.png
Biotronik: About
btk ser.png
Biotronik: About

V pace suppression

Opens ups the Av delay by 450ms to search for intrinsic V. If 6 consecutive Vs occur, the mode will switch from DDD – ADI. If 3 of 8 beats are Vp, mode will switch back to DDD, or 2 in a row .


Cannot use CLS with Vp suppression


Search occurs 30 sec > 1 min > 2 min > 4 min > .... > 128 min > 20

btk mvp.png
Biotronik: About

Accessed from Parameters; Bradycardia; Mode

•Mode: Select the mode [DDD-ADI] or [DDDR-ADIR]

•V pace suppression: OFF; on

•Pacing suppression after consecutive V sense:

1 ... (1) ... 8, 6

•Pacing supported after X-out-of-8-cycles: 1, 2, 3, 4

mvp 3.png
Biotronik: About
mvp btk 4.jpg
Biotronik: About

Hysteresis

Accessed from Parameters; Bradycardia; basic rate/night rate; hysteresis

• hysteresis (ppm):  OFF; -5 to -

•repetitive cycles:  OFF; 1 ... (1) ... 15

•scan cycles:  OFF; 1 ... (1) ... 15

btk hyst1.png
Biotronik: About
btk hyst2.jpg
Biotronik: About
Biotronik: About
bottom of page