Biotronik
Summary
Single and dual chamber:
•Hysteresis with AV search for dual chamber
Dual chamber:
•AV hysteresis and IRS+
•Ventricular pace suppression
AV search
Positive AV hysteresis. After 180 Vp beats -Opens up the AV delay by x ms for 5 cycles (AV scan cycle).
When an intrinsic R wave is detected, the AV delay is extended x ms for 5 cycles (AV repetitive AV hysteresis).


Accessed from Parameters; Bradycardia; Dynamic AV delay; AV hysteresis
• AV hysteresis (ppm): OFF; Negative, low 250ms; medium 285ms; high 325ms; IRSplus 400ms
•AV repetitive hysteresis (cycles): OFF; 1 ... (1) ... 10
•AV scan hysteresis (cycles): OFF; 1 ... (1) ... 10
​


V pace suppression
Opens ups the Av delay by 450ms to search for intrinsic V. If 6 consecutive Vs occur, the mode will switch from DDD – ADI. If 3 of 8 beats are Vp, mode will switch back to DDD, or 2 in a row .
Cannot use CLS with Vp suppression
Search occurs 30 sec > 1 min > 2 min > 4 min > .... > 128 min > 20

Accessed from Parameters; Bradycardia; Mode
•Mode: Select the mode [DDD-ADI] or [DDDR-ADIR]
•V pace suppression: OFF; on
•Pacing suppression after consecutive V sense:
1 ... (1) ... 8, 6
•Pacing supported after X-out-of-8-cycles: 1, 2, 3, 4


Hysteresis
Accessed from Parameters; Bradycardia; basic rate/night rate; hysteresis
• hysteresis (ppm): OFF; -5 to -
•repetitive cycles: OFF; 1 ... (1) ... 15
•scan cycles: OFF; 1 ... (1) ... 15