Boston Scientfic
Summary
Single and dual chamber:
•Hysteresis with AV search for dual chamber
​
Dual chamber:
•Rhythm IQ
Rhythm IQ
During normal conduction, RHYTMIQ operates like 2 pacemakers running concurrently – one in AAI(R) mode and one is VVI mode
​
The device provides AAI(R) at the LRL and/or sensor indicated rate with backup VVI pacing at a rate 15 bpm slower than the LRL
​
Mode is switched to DDD if 3/11 ventricular paced beats are detected
AV search + is used to search for intrinsic conduction to switch back to AAI – switches when <2/10 cycles are ventricular paced
​
Accessed from brady parameters; RHYTHMIQ
•RHYTHMIQ: AAI with VVI backup, Off
​

AV search +
Accessed from brady parameters; timing/rate enhancements/magnet/noise; timing
•AV search +: on, Off
•Search AV delay 30-400 ms (10ms increments)
•Search interval 32, 64, 128, 256, 512, 1024 cycles
​




Hysteresis
Accessed from brady parameters; Rate enhancements; Rate hysteresis;
•Hysteresis Offset (bpm): off, -5 - -80 (5 bpm increments)
•Search Hysteresis: off, 256-4096 cycles (powers of 2)