-
81 year old female with dual chamber implanted for symptomatic sinus pauses
-
Currently programmed AAIR-DDDR 60-130
-
Underlying sinus rhythm 1st degree AV block
-
Initial appearance of markers appears to be regular atrial rate at a rate of ~160bpm
-
Close inspection of EGMs shows sinus rhythm with 1st degree and FFRW
-
No true AF seen
​​​​​​​​​​​​​​​​​​
What is FFRW (Far Field R Wave)
-
common issue where atrial lead sense a far field ventricular signal
-
It is more likely to occur in the following scenarios
unipolar le-unipolar lead sensing -
atrial lead position close to the tricuspid valve
-
very sensitive programming
-
wide electrode spacing​​​​​​​​​​​​​​​
​
Inappropriate mode switches
-
records EGMs for episodes which may write over true arrhythmia episodes and also use extra battery
-
inappropriate ventricular pacing inhibition due to being a non tracking mode
--may result in AV dysnchrony due to AV block
- may result in withholding required biventricular pacing
-may incorrectly bin tachycardias in an ICD (potential for A>V rather than 1:1 bin) -
​
-
Adjust atrial sensitivity? -
Currently 0.30mV ( p wave sensing is 2mV) - risk of under-sensing true atrial rhythms
​​​
-
Increase/adjust PBAB?
-
Increase PVAB, so that AR falls into blanking period.
​
MEDTRONIC DEVICES BEHAVE DIFFERENTLY
-
Technically, in this example the FFRW is blanked (not in refractory).
-
To eliminate the ABlank marker, PVAB can be adjusted to Partial + (described below) or Absolute which would behave similarly to other device brands
-
​
-
-Partial: 30ms of absolute blanking followed by the rest of the PVAB (e.g., PVAB 150, (150-30 = 120)) influencing mode switch counters (but not timing or NCAP, PMT, etc)
-Partial+: 30ms of absolute blanking followed by the rest of the PVAB but at a reduced sensitivity (e.g., PVAB 150, (150-30 = 120)). Events in this window will still influence AMS and ICD discrimination algorithms.
-Absolute: standard PVAB programming​​​​​​​