top of page

Background

  • 81 year old female with dual chamber implanted for symptomatic sinus pauses

  • Currently programmed AAIR-DDDR 60-130

  • Underlying sinus rhythm 1st degree AV block

 

Routine Clinic check

3 AF episodes (<1% burden)

​

What does this episode show?

​​

 

Findings

  • Initial appearance of markers appears to be regular atrial rate at a rate of ~160bpm

  • Close inspection of EGMs shows sinus rhythm with 1st degree and FFRW

  • No true AF seen

​​​​​​​​​​​​​​​​​​

What is FFRW (Far Field R Wave)
 

  • common issue where atrial lead sense a far field ventricular signal

  • It is more likely to occur in the following scenarios
    unipolar le-unipolar lead sensing

  • atrial lead position close to the tricuspid valve

  • very sensitive programming

  • wide electrode spacing​​​​​​​​​​​​​​​

​

Implications 
 

Inappropriate mode switches

  • records EGMs for episodes which may write over true arrhythmia episodes and also use extra battery

  • inappropriate ventricular pacing inhibition due to being a non tracking mode
    --may result in AV dysnchrony due to AV block
    - may result in withholding required biventricular pacing
    -may incorrectly bin tachycardias in an ICD (potential for A>V rather than 1:1 bin)

  • ​

Troubleshooting


  • Adjust atrial sensitivity?

  • Currently 0.30mV ( p wave sensing is 2mV) - risk of under-sensing true atrial rhythms

​​​

  • Increase/adjust PBAB?

  • Increase PVAB, so that AR falls into blanking period.

​

MEDTRONIC DEVICES BEHAVE DIFFERENTLY

  • Technically, in this example the FFRW is blanked (not in refractory).

  • To eliminate the ABlank marker, PVAB can be adjusted to Partial + (described below) or Absolute which would behave similarly to other device brands

  • ​

  • -Partial: 30ms of absolute blanking followed by the rest of the PVAB (e.g., PVAB 150, (150-30 = 120)) influencing mode switch counters (but not timing or NCAP, PMT, etc)
    -Partial+: 30ms of absolute blanking followed by the rest of the PVAB but at a reduced sensitivity (e.g., PVAB 150, (150-30 = 120)). Events in this window will still influence AMS and ICD discrimination algorithms.
    -Absolute: standard PVAB programming
    ​​​​​​​

bottom of page