Non responders


Phrenic stimulation


Pro-arrhythmia

Complications of LV implants: About

Non responders

Non responders may be due to multiple factors

​

Low pacing percentage (<90%)

​

Loss of AV synchrony

​

BiV pacing inhibition

​

LV capture issues

​

RV capture issues

​

Other

Complications of LV implants: About
lowp.png
Complications of LV implants: About
av.png
Complications of LV implants: About
biv.png
Complications of LV implants: About
djs.png
Complications of LV implants: About
rv.png
Complications of LV implants: About

Phrenic stimulation

Decrease pacing output


Reprogramming lead from unipolar to bipolar


Changing LV pacing vector

​

Repositioning lead


Turning OFF auto-capture algorithms for LV lead


Nothing done if only occurs in rare postural positions

​

Reassurance to patient in important

​

Complications of LV implants: About

Proarrhythmia

LV pacing inducing VT


Vary rare and not well documented


Can be site dependent

Complications of LV implants: About
pro.png
Complications of LV implants: About
Complications of LV implants: About