top of page
Complications of CRT
Complications of LV implants: Welcome
Non responders
Phrenic stimulation
Pro-arrhythmia
Complications of LV implants: About
Non responders
Non responders may be due to multiple factors
​
Low pacing percentage (<90%)
​
Loss of AV synchrony
​
BiV pacing inhibition
​
LV capture issues
​
RV capture issues
​
Other
Complications of LV implants: About

Complications of LV implants: About

Complications of LV implants: About

Complications of LV implants: About

Complications of LV implants: About

Complications of LV implants: About
Phrenic stimulation
Decrease pacing output
Reprogramming lead from unipolar to bipolar
Changing LV pacing vector
​
Repositioning lead
Turning OFF auto-capture algorithms for LV lead
Nothing done if only occurs in rare postural positions
​
Reassurance to patient in important
​
Complications of LV implants: About
Proarrhythmia
LV pacing inducing VT
Vary rare and not well documented
Can be site dependent
Complications of LV implants: About

Complications of LV implants: About
Complications of LV implants: About
bottom of page