Redetection refers to determining if the tachyarrhytmia persists after therapy has been delivered.

Abbott

Redetection compares intervals in the same manner as detection

abott redetect.jpg

Biotronik

Redetection compares intervals in the same manner as detection

Boston Scientific

Redetection looks for 8/10 fast intervals to occur.


A redetection duration period is then set in seconds.


As long as 6/10 intervals remains fast, the arrhythmia is determined as sustained.

​

bsc redetect.png

Boston Scientific - Subcutaneous ICD

Same as detection but 14/24 beats

​

Medtronic

Compares intervals in the same manner as detection

​

Microport

After a therapy, persistence counter is not used (majority counter only – 6/8)

​