End of episode refers to determining if the VA episode has terminated (either spontaneously or after therapy).

Abbott

Return to sinus rhythm intervals are programmable

•Short: 3 intervals

•Normal: 5 intervals

•Long: 7 intervals

abbott end of episode.jpg

Biotronik

When 12/16 intervals (non programmable) are slower than the slowest VT zone

bio end of episode.png

Boston Scientific

Redetection looks for 8/10 fast intervals to occur.


An end of episode duration period is then set in seconds (non programmable)

•Non treated episodes  10 seconds

•ATP only therapy  10 seconds

•Shock therapy  30 seconds


As long as 6/10 intervals remains fast, the arrhythmia is determined as sustained.

​

​

bsc end of.png

Boston Scientific - Subcutaneous ICD

Once in an episode, the device applies a hysteresis of 40ms under the lowest therapy zone.

​

24 intervals need to be below this rate to declare end of episode

​

sub cut end.png

Medtronic

8 consecutive VS intervals

mdt end.png

Microport

When at least 75% of the last 8 cycles are slower than the tachycardia zones