top of page

Summary of algorithms

Medtronic

•Rate Drop response (RDR)


Boston

•Sudden brady response (SBR)

Microport

•Dplus mode


Biotronik

•Rate Fading

•Closed loop stimulation (CLS)

Abbott

•Advanced hysteresis

Any

•Hysteresis

Device algorithms: About

Medtronic - rate drop response

Rate Drop Response (RDR) is intended to provide backup pacing and prevent associated symptoms in patients who experience occasional episodes of significant drop in heart rate (e.g., syncope from cardioinhibitory and mixed forms of carotid sinus syndrome)

There are two types of rate drop detection:

•Drop Detect occurs when there is a Sudden Rate Drop below the Drop Size and inside the Detection Window

•Low Rate Detect occurs when there is a Sudden Rate Drop to the lower rate.

Device algorithms: About

Drop detect

Drop Detect occurs when there is a Sudden Rate Drop below the Drop Size &inside the Detection Window

vv8.png
Device algorithms: About

Low rate detect

Low Rate Detect occurs when there is a Sudden Rate Drop to the lower rate.

vv9.png
Device algorithms: About

Medtronic programming

m1.png
Device algorithms: About
m2.png
Device algorithms: About
m3.png
Device algorithms: About

Accessed from parameters; additional features; Rate therapies


•Detection Type:    low rate, drop, both, off 


•Intervention Rate:  60, 70, 75, ...100, ...180 bpm   (not 65 or 85 (magnet rates)


•Intervention Duration:  1, 2, 3, ...15 minutes


•(LR)Detection beats:  1, 2, 3 beats 


•(D)Drop Rate:  30, 40, 50, ...100 bpm


•(D)Drop Size:  10, 15, 20, 25, ...50 bpm


•(D)Detection Window:  10, 15, 20, 25, ...30 s; 1, 1.5, 2,   2.5 minutes

Device algorithms: About

Boston - Sudden brady response

SBR is designed to respond to sudden decreases in intrinsic atrial rates by applying dual-chamber pacing at an elevated rate (DDD(R) modes).


SBR is declared when the atrial chamber has been continuously sensed for a programmable time, followed by a sudden decrease in atrial rate such that atrial pacing occurs at the lower rate limit or the sensor indicated rate for a programmable number of cycles.


The decrease in atrial rate preceding the paced events must exceed 10 bpm.


The SBR algorithm continually monitors the average of the atrial rate and this average is updated each cardiac cycle.

SBR Detect Time: the programmable time interval during which the atrium must be 100% sensed.


SBR Number of Beats: criteria are applied once the decrease in atrial rate has been detected and pacing occurs (either LRL or sensor indicated rate).


Atrial pacing must occur for the programmable number of consecutive intervals before the SBR criteria are met.


SBR Therapy Rate Offset: calculated by using the patients average atrial rate before the bradycardia and adding a programmable positive offset.


SBR therapy duration: is the programmable time interval during which SBR pacing therapy will be used.


Once pacing therapy has been delivered, the atrial pacing rate will be decreased using a 12% rate smoothing down factor until the lower rate or sensor indicated rate is reached.

sbr1.png
Device algorithms: About

Boston Programming

sbr2.png
Device algorithms: About
sbr3.png
Device algorithms: About
Device algorithms: About
sbr4.png

Accessed from brady parameters; rate enhancements


•Sudden Brady Response:  off, on

•SBR Detect Time:    1-5 minutes (1 min increments) (5 min) 

•SBR number of beats:  1-8 cycles (1 increments) (4 cycles)

•SBR therapy rate offset:  5-40 bpm (5 bpm increments) (5 bpm)

•SBR therapy duration:  1-15 minutes (1 min increments) (10 mins)

Device algorithms: About

Microport D Plus mode

Dplus is an automatic AV delay histeresis algorithm that extends the AV delay (pseudo AAI) to encourage intrinsic rhythm but shortens the AV delay when AV delay worsens (DDD).

A switch fom DDD to pseudo AAI occurs after 8 consecutive cycles of AV conduction

Device algorithms: About
dplus4.png
Device algorithms: About
Device algorithms: About

Accessed from basic parameters; mode

•Mode: Dplus, Dplus-R, Dplus/DDIR

•Activation of Dplus results is automatic AV delay calculation

dplus.png
Device algorithms: About
Device algorithms: About

DPlus pacing mode with acceleration algorithm

•in order to access to acceleration function, program rate hysteresis with a value differnt from 0%

dplus2.png
Device algorithms: About
Device algorithms: About

DPlus pacing mode with AV delay shortening

•in order to access AV delay shortening function, program acceleration with a value different from 0%

dplus3.png
Device algorithms: About

Biotronik - Rate fading

Prevent a sudden drop in heart rate when the pulse generator detects an abrupt decrease in the intrinsic rate.


Rate fading can be used to prevent potential reactions such as dizziness, light headedness, lack of energy and fainting.  


Calculates the fading rate, which is a 4 beat average of the intrinsic rate reduced by 10 ppm.


When the intrinsic rate drops considerably (below the fading rate), the pacing rate begins at the rate fading rate and then decreases gradually by the programmable decay rate to the sensor or base rate.

•Rate increase: the maximum rate of change in the pacing rate if the sensor signal indicates increasing exertion.


•Rate decrease: the maximum rate of change in the pacing rate if the sensor signal indicates decreasing exertion.

rf1.png
Device algorithms: About
rf2.png
Device algorithms: About
rf3.png
Device algorithms: About

Biotronik programming

rf5.png
Device algorithms: About
rf4.png
Device algorithms: About

Accessed from bradycardia; sensor/rate fading


•Rate Fading:  off, on 


•Rate Increase:  1-10 bpm (1 bpm increments) (4)


•Rate decrease:  0.1, 0.2, 0.5, 1.0  bpm

Device algorithms: About

Hysteresis

Hysteresis may be set up in a way to mimic behavior of rate drop algorithms

This can be done by programming a

•high base rate, e.g., 100 ppm

•with a low hysteresis rate e.g., 50 ppm


Therefore, when the sensed rate drops to the hysteresis rate, fast pacing at the lower rate will occur

hys.png
Device algorithms: About

Abbott - advanced hysteresis

Advanced hysteresis may be set up in a way to mimic behavior of rate drop algorithms using the hysteresis rate as the drop rate, and an intervention rate with intervention duration and recovery time to aid the process.

Device algorithms: About

Abbott programming

hysttttt.png
Device algorithms: About
hysss.png
Device algorithms: About

Accessed from bradycardia; rate/timing

•Hysteresis rate (bpm):  off; 30-150 in steps of 5

•Cycle Count: 1-16 (in steps of 1)

•Intervention rate (bpm):  off, 80-120 in steps of 10; intrinsic +0, intrinsic +10; intrinsic +20; intrinsic +30; same as base rate 

•Intervention duration (bpm):  1-10 bpm (1 bpm increments)

•Recovery time:  fast; medium; slow; very slow

Device algorithms: About

Limitations of rate drop algorithms

The drop in the heart rate response for a vasovagal episode occurs in the 3rd stage of the episode and usually when the blood pressure response has already begun to dip.

Thus having algorithms that respond to the change in heart rate may not be effective in preventing the episode of syncope.

vvs.png
Device algorithms: About

CLS measures the intracardiac impedance throughout EACH ventricular contraction.


During the “resting state” (monitored by accelerometer) the impedance values will establish a continuously updated reference curve, which reflects the contraction dynamics at rest.

cls1.png
Device algorithms: About

Biotronik - Programming

cls2.png
Device algorithms: About

Accessed from Parameters; Bradycardia


•Mode:   [DDD-CLS] or [VVI-CLS]


•[CLS] CLS rate:   85% of 100% max HR


•[CLS- Expert Options]

•CLS Response (slope):   very low, low, medium, high, very high

•Resting Rate Control:   off, +10, +20, +30, +40, +50

•V pace required:   yes, no

For syncopal patients, program resting rate to off. Thus CLS will increase to max HR. Moreover, program CLS response to high so that these patients get a quick response.

Device algorithms: About
Device algorithms: About
bottom of page